Could posterior-anterior projection cervical spine radiographs improve image quality and dose reduction

Journal article


Faulkner, R. and Lockwood, P. 2022. Could posterior-anterior projection cervical spine radiographs improve image quality and dose reduction . Radiography Open. 8 (1). https://doi.org/10.7577/radopen.5004
AuthorsFaulkner, R. and Lockwood, P.
Abstract

Introduction: Anterior-posterior (AP) cervical spine X-rays are routine examinations to assess degenerative change, persistent pain and traumatic injuries. Multiple radiosensitive organs lie anteriorly within this anatomical region, increasing the stochastic risk of cancer. If a posterior-anterior (PA) projection was utilised, the radiation dose could potentially be reduced. The hypothesis of this study is to evaluate the change in radiation dose and image quality between AP and PA positions.

Materials and Methods: An anthropomorphic phantom was positioned AP erect against a digital radiography (DR) detector with 30 thermoluminescent dosimeters (TLDs) inserted to record the thyroid, breast, ovaries, and testes absorbed radiation dose at an exposure of 66 kV and 8 mAs. The phantom was repositioned PA erect and repeated. Images were assessed against an image quality criteria Likert scale by qualified radiographers. The mean and standard deviations were calculated for dose and image quality and compared using a t-test and Wilcoxon Signed Ranks Test.

Results: The PA erect cervical spine reduced radiation dose to the right thyroid by 92% (44.7 µGy; p=0.00) and the left thyroid by 89% (43.7 µGy; p=0.00), with further reductions in scatter dose to the breasts (0.35-0.45 µGy; p=0.85), ovaries (0.41 µGy; p=0.57), and testes (0.04 µGy; p=0.98). Image quality scores for the end plates, pedicles, joint spaces, spinous and transverse processes, cortical and trabecular bone patterns, and soft tissues were near equivalent (p=0.32).

Conclusion: Data analysis suggests that PA cervical spine positioning for X-rays in the laboratory adheres to as low as reasonably practicable (ALARP) guidance on X-ray examinations to reduce radiation dose to male and female internal organs (thyroid, breast, ovaries) without a reduction in image quality compared to AP positioning. Further research in clinical practice is advised.

KeywordsCervical spine; Image quality; Radiation dose; Scatter radiation; Phantom study
Year2022
JournalRadiography Open
Journal citation8 (1)
PublisherOslo Met
ISSN2387-3345
Digital Object Identifier (DOI)https://doi.org/10.7577/radopen.5004
Official URLhttps://journals.oslomet.no/index.php/radopen/article/view/5004
Publication dates
Print11 Nov 2022
Publication process dates
Accepted01 Nov 2022
Deposited04 Jan 2023
Publisher's version
License
File Access Level
Open
Output statusPublished
References

1. NHS England. Diagnostic Imaging Dataset Annual Statistical Release 2019/2020. Diagnostic Imaging Dataset Annual Statistical Release 2019/2020. Available from: https://www.england.nhs.uk/statistics/wp-content/uploads/sites/2/202... [accessed October 4, 2022].
2. Murphy A. Cervical Spine (AP View). Radiopaedia. Available from: https://radiopaedia.org/articles/cervical-spine-ap-view?lang=gb [accessed October 4, 2022].
3. Ben-Shlomo A., Bartal G., Mosseri M., Avraham B., Leitner Y., Shabat S. Effective dose reduction in spine radiographic imaging by choosing the less radiation-sensitive side of the body. The Spine Journal 2016;16(4):558–63. Doi: 10.1016/j.spinee.2015.12.012.
4. Cancer Research UK. Cancer Statistics for the UK. Cancer Statistics for the UK. Available from: https://www.cancerresearchuk.org/health-professional/cancer-statisti... [accessed October 4, 2022].
5. Chan CTP., Fung KKL. Dose optimization in pelvic radiography by air gap method on CR and DR systems – A phantom study. Radiography 2015;21(3):214–23. Doi: 10.1016/j.radi.2014.11.005.
6. UK Government. Ionising Radiation (Medical Exposure) Regulations (IR(ME)R) 2017 (SI 2017/1322), London: HMSO; 2017.
7. Majeed A., Kanwal A., Naz N. Diagnostic Accuracy of Plain X-ray in Diagnosis of Cervical Spine Fracture, Keeping CT as Gold Standard. Institute of Medical Science 2016;12(3):171–4.
8. Whitley S., Sloane C., Jefferson G., Holmes K., Anderson C. Clarke’s Pocket Handbook for Radiographers , 13th ed., London: Hodder Arnold; 2010.
9. Davey E., England A. AP versus PA positioning in lumbar spine computed radiography: Image quality and individual organ doses. Radiography 2015;21(2):188–96. Doi: 10.1016/j.radi.2014.11.003.
10. Holm T., Palmer PES., Lehtinen E. World Health Organization. Basic Radiological System: Manual of Radiographic Technique, Geneva: World Health Organization; 1986.
11. Slosar P. Cervical Spine Anatomy. Spine Health . Available from: https://www.spine-health.com/conditions/spine-anatomy/cervical-spine... [accessed October 4, 2022].
12. Holmes K., Elkington M., Harris P. Clark’s Essential Physics In Imaging For Radiographers. , Lancaster: University of Cumbria, School of Medical Imaging Sciences ; 2013.
13. Whitley S., Sloane C., Hoadley G., Moore A., Alsop C. Clarke’s Pocket Handbook for Radiographers, 12th ed., London: Hodder Arnold; 2005.
14. UK Government. The Ionising Radiation (Medical Exposure) Regulations 2017, London: UK Statutory Instruments; 2017.
15. Landauer. ⅛" x ⅛" x 0.15" TLD-100H . TLD Chip: Single Point Radiation Assessments. Available from: https://www.landauer.co.uk/produit/tld-chip-single-point-radiation-a... [accessed October 4, 2022].
16. J. Stratakis AP. Chapter 1 Dosimetry. Radiation Dose Management of Pregnant Patients, Pregnant Staff and Paediatric Patients, IOP Publishing; 2019.
17. Thermo Electron Corporation. Thermo Fisher Scientific Harshaw TLD Model 5500 Reader with WinREMS Operator’s Manual (5500-W-O-0805-006). , 2005.
18. Hanna DW. Development and optimization of a thermoluminescent dosimeter (TLD) analyser system for low-dose measurements utilizing photon counting techniques. Kansas State University, Kansas , 1979.
19. International Commission on Radiological Protection. The 2007 Recommendations of the International Commission on Radiological Protection. Annals of the ICRP 2007;103.
20. European Guidelines on Quality Criteria for Diagnostic Radiographic Images. European Guidelines on Quality Criteria for Diagnostic Radiographic Images, 1996.
21. Hedberg EC., Ayers S. The power of a paired t-test with a covariate. Soc Sci Res 2015;50:277–91. Doi: 10.1016/j.ssresearch.2014.12.004.
22. The British Institute of Radiology. Guidance on using shielding on patients for diagnostic radiology applications, London; 2020.
23. Cancer Research UK. Risks and Causes of Thyroid Cancer. Risks and Causes of Thyroid Cancer. Available from: https://www.cancerresearchuk.org/about-cancer/thyroid-cancer/causes-... [accessed October 4, 2022].
24. Robinson JB., Ali RM., Tootell AK., Hogg P. Does collimation affect patient dose in antero-posterior thoraco-lumbar spine? Radiography 2017;23(3):211–5. Doi: 10.1016/j.radi.2017.03.012.
25. Green C., Karnati G., Thomson K., Subramanian A. Lumbar spine radiographs — is it time for widespread adoption of posteroanterior projection? Br J Radiol 2019;92(1103):20190386. Doi: 10.1259/bjr.20190386.
26. Health and Care Professions Council. Standards of Proficiency - Radiographers, London; 2013.
27. Stephenson‐Smith B., Neep MJ., Rowntree P. Digital radiography reject analysis of examinations with multiple rejects: an Australian emergency imaging department clinical audit. J Med Radiat Sci 2021;68(3):245–52. Doi: 10.1002/jmrs.468.
28. Don S., MacDougall R., Strauss K., Moore QT., Goske MJ., Cohen M., et al. Image Gently Campaign Back to Basics Initiative: Ten Steps to Help Manage Radiation Dose in Pediatric Digital Radiography. American Journal of Roentgenology 2013;200(5):W431–6. Doi: 10.2214/AJR.12.9895.
29. Singh T., Muscroft N., Collier N., England A. A comparison of effective dose and risk for different collimation options used in AP shoulder radiography. Radiography 2022;28(2):394–9. Doi: 10.1016/j.radi.2021.11.007.
30. Johnston J., Fauber TL. Essentials of radiographic physics and imaging, Elsevier Health Sciences; 2015.
31. National Institute for Health and Care Excellence. Spinal injury: assessments and initial management, London; 2016.
32. Head Injury: Assessment and Early Management. National Institute for Health and Care , London; 2021.
33. Loughman E., Rowan M., Kenny P. Does magnification increase total unsharpness in DR radiography? Physica Medica 2017;42:353. Doi: 10.1016/j.ejmp.2017.05.007.
34. Chaparian A., Kanani A., Baghbanian M. Reduction of radiation risks in patients undergoing some X-ray examinations by using optimal projections: A Monte Carlo program-based mathematical calculation. J Med Phys 2014;39(1):32. Doi: 10.4103/0971-6203.125500.
35. Bell D. Radiographic Distortion. Radiopaedia. Available from: https://radiopaedia.org/articles/radiographic-distortion?lang=gb [accessed October 4, 2022].
36. Nunn H. The Cervical Spine. Image Interpretation . Available from: https://www.imageinterpretation.co.uk/cervical.php [accessed October 4, 2022].

Permalink -

https://repository.canterbury.ac.uk/item/936v3/could-posterior-anterior-projection-cervical-spine-radiographs-improve-image-quality-and-dose-reduction

Download files


Publisher's version
5004-ny-template.pdf
License: CC BY 4.0
File access level: Open

  • 182
    total views
  • 123
    total downloads
  • 17
    views this month
  • 16
    downloads this month

Export as

Related outputs

A survey of the NHS reporting radiographer workforce in England
Lockwood, P., Burton, C., Shaw, T. and Woznitza, N. 2024. A survey of the NHS reporting radiographer workforce in England. Radiography Open. 10 (1), pp. 1-18. https://doi.org/10.7577/radopen.5635
Comparing the standard knee X-ray exposure factor, 10 kV rule, and modified 10 kV rule techniques in digital radiography to reduce patient radiation dose without loss of image quality
Lockwood, P. and Wenman, A. 2024. Comparing the standard knee X-ray exposure factor, 10 kV rule, and modified 10 kV rule techniques in digital radiography to reduce patient radiation dose without loss of image quality. Radiography. 30 (2), pp. 574-581. https://doi.org/10.1016/j.radi.2024.01.013
Obesity bias in diagnostic radiography students: A survey of attitudes, perceptions and technical confidence
Tamburrini, N. and Lockwood, P. 2023. Obesity bias in diagnostic radiography students: A survey of attitudes, perceptions and technical confidence. Radiography. 30 (1), pp. 202-208. https://doi.org/10.1016/j.radi.2023.11.006
Care of transgender patients by diagnostic radiographers: What can be learnt from the literature
Hammond, C and Lockwood, P. 2023. Care of transgender patients by diagnostic radiographers: What can be learnt from the literature. Radiography. 30 (1), pp. 145-150. https://doi.org/10.1016/j.radi.2023.10.020
Assessing the barriers and enablers to the implementation of the diagnostic radiographer musculoskeletal X‑ray reporting service within the NHS in England: a systematic literature review
Lockwood, P., Burton, C., Woznitza, N. and Shaw, T. 2023. Assessing the barriers and enablers to the implementation of the diagnostic radiographer musculoskeletal X‑ray reporting service within the NHS in England: a systematic literature review. BMC Health Services Research. 23 (1270), pp. 1-41. https://doi.org/10.1186/s12913-023-10161-y
Impostor phenomenon traits in radiography students: findings from a UK pilot survey
Gibson, C. and Lockwood, P. 2023. Impostor phenomenon traits in radiography students: findings from a UK pilot survey . Radiography. 30 (1), pp. 61-65. https://doi.org/10.1016/j.radi.2023.10.005
Investigating the adjacent patient radiation dose received during a simulated ward chest X-ray examination
Langfield, H and Lockwood, P. 2023. Investigating the adjacent patient radiation dose received during a simulated ward chest X-ray examination. Radiography Open. 9 (1), pp. 1-14. https://doi.org/10.7577/radopen.5354
Research ethics applications: Back to basics: What you ought to know about research ethics applications
Lockwood, P. 2023. Research ethics applications: Back to basics: What you ought to know about research ethics applications. Insights into Imaging. Spring (7), pp. 39-45.
An investigation into the clinical scope of practice of MRI reporting radiographers within the United Kingdom
Mitchell, M. and Lockwood, P. 2023. An investigation into the clinical scope of practice of MRI reporting radiographers within the United Kingdom. Radiography. 29 (3), pp. 489-495. https://doi.org/10.1016/j.radi.2023.02.015
Assessing the barriers and enablers to the diagnostic radiographer X-ray reporting service within the NHS in England: A systematic literature review
Lockwood, P. 2022. Assessing the barriers and enablers to the diagnostic radiographer X-ray reporting service within the NHS in England: A systematic literature review. York University.
Lead-rubber shielding effect on radiation dose to the gonads from a bilateral hand X-ray examination
Welborn, D. and Lockwood, P. 2022. Lead-rubber shielding effect on radiation dose to the gonads from a bilateral hand X-ray examination. Radiography. 28 (2), pp. 360-365. https://doi.org/10.1016/j.radi.2021.12.013
Diagnostic radiography students' perceptions towards communication with service users who are deaf or hearing impaired.
Nolan-Bryant, A. and Lockwood, P. 2022. Diagnostic radiography students' perceptions towards communication with service users who are deaf or hearing impaired. Radiography. 29 (1), pp. 207-214. https://doi.org/10.1016/j.radi.2022.11.008
A cross-sectional student survey of the impact of the Covid-19 lockdowns on clinical placement in England
Hinds, Z. and Lockwood, P. 2022. A cross-sectional student survey of the impact of the Covid-19 lockdowns on clinical placement in England. Radiography. 29 (1), pp. 190-199. https://doi.org/10.1016/j.radi.2022.11.006
Pre-registration UK diagnostic radiography student ability and confidence in interpretation of chest X-rays
Lockwood, P. and Khan, A. 2021. Pre-registration UK diagnostic radiography student ability and confidence in interpretation of chest X-rays. Radiography Open. 7 (1), pp. 1-13. https://doi.org/10.7577/radopen.4529
Efficacy, utility, and validity in Computed Tomography head reporting by radiographers
Lockwood, P. 2021. Efficacy, utility, and validity in Computed Tomography head reporting by radiographers. PhD Thesis Canterbury Christ Church University School of Allied and Public Health Professions
COVID-19: A literature review of the impact on diagnostic radiography students
Astirbadi, D. and Lockwood, P. 2021. COVID-19: A literature review of the impact on diagnostic radiography students. Radiography. https://doi.org/10.1016/j.radi.2021.09.009
Multi-professional image interpretation: performance in preliminary clinical evaluation of appendicular radiographs
Lockwood, P. and Pittock, L. 2019. Multi-professional image interpretation: performance in preliminary clinical evaluation of appendicular radiographs. Radiography. https://doi.org/10.1016/j.radi.2019.04.013
Computed tomography head and facial bones review of a 2,700 year old Egyptian mummy
Lockwood, P., Elliott, J., Nelson, A. and Harris, S. 2019. Computed tomography head and facial bones review of a 2,700 year old Egyptian mummy . BJR Case Reports. https://doi.org/10.1259/bjrcr.20190076
Visual function assessment of diagnostic radiography students
Lockwood, P. and Blackman, A. 2019. Visual function assessment of diagnostic radiography students. Radiography. https://doi.org/10.1016/j.radi.2019.10.012
An evaluation of CT head reporting radiographers' scope of practice within the United Kingdom
Lockwood, P. 2019. An evaluation of CT head reporting radiographers' scope of practice within the United Kingdom. Radiography. 26 (2), pp. 102-109. https://doi.org/10.1016/j.radi.2019.09.001
Image Interpretation by radiographers in brain, spine and knee MRI examinations: Findings from an accredited postgraduate module
Lockwood, P. and Dolbear, G. 2018. Image Interpretation by radiographers in brain, spine and knee MRI examinations: Findings from an accredited postgraduate module. Radiography. https://doi.org/10.1016/j.radi.2018.05.009
Evaluation of an equilibrium phase free-breathing dynamic contrast-enhanced MRI prototype sequence compared to traditional breath-held MRI acquisition in liver oncology patients
Hopkinson, G., Lockwood, P. and Dolbear, G. 2018. Evaluation of an equilibrium phase free-breathing dynamic contrast-enhanced MRI prototype sequence compared to traditional breath-held MRI acquisition in liver oncology patients. Radiography. https://doi.org/10.1016/j.radi.2018.01.001
Nuclear medicine image interpretation by radiographers: findings of an accredited postgraduate module
Lockwood, P. and Dolbear, G. 2018. Nuclear medicine image interpretation by radiographers: findings of an accredited postgraduate module. Radiography. https://doi.org/10.1016/j.radi.2018.11.007
CT Sinus and facial bones reporting by radiographers: findings of an accredited postgraduate programme
Lockwood, P. 2017. CT Sinus and facial bones reporting by radiographers: findings of an accredited postgraduate programme. Dentomaxillofacial Radiology. 46 (4). https://doi.org/10.1259/dmfr.20160440
Exploring variation and trends in adherence to national occupational standards for reporting radiographers
Lockwood, P. 2017. Exploring variation and trends in adherence to national occupational standards for reporting radiographers. Journal of Social Science & Allied Health Professions. 1 (1), pp. 20-27.
Observer performance in Computed Tomography head reporting
Lockwood, P. 2017. Observer performance in Computed Tomography head reporting. Journal of Medical Imaging and Radiation Sciences. 48 (1), pp. 22-29. https://doi.org/10.1016/j.jmir.2016.08.001
Exploring the benefits of magnetic resonance imaging reporting by radiographers: A UK perspective
Lockwood, P. 2016. Exploring the benefits of magnetic resonance imaging reporting by radiographers: A UK perspective. Journal of Medical Imaging and Radiation Sciences. 47 (2). https://doi.org/10.1016/j.jmir.2015.12.083
An economic evaluation of introducing a skills mix approach to CT head reporting in clinical practice.
Lockwood, P. 2016. An economic evaluation of introducing a skills mix approach to CT head reporting in clinical practice. Radiography. https://doi.org/10.1016/j.radi.2015.09.004
Intraorbital foreign body detection and localisation by radiographers: a preliminary JAFROC observer performance study
Lockwood, P., Pittock, L., Lockwood, C., Jeffery, C. and Piper, K. 2015. Intraorbital foreign body detection and localisation by radiographers: a preliminary JAFROC observer performance study. Radiography. 2015, pp. 1-5. https://doi.org/10.1016/j.radi.2015.06.005
Out of the comfort zone (Part 2)
Lockwood, P. 2013. Out of the comfort zone (Part 2).
AFROC analysis of reporting radiographer’s performance in CT head interpretation
Lockwood, P. and Piper, K. 2015. AFROC analysis of reporting radiographer’s performance in CT head interpretation. Radiography. 21 (3), pp. e90-e95. https://doi.org/10.1016/j.radi.2015.04.001
Out of the comfort zone (Part 1)
Lockwood, P. 2013. Out of the comfort zone (Part 1).
Origins of the Reporting Radiographer
Lockwood, P. 2013. Origins of the Reporting Radiographer.
Patient safety and quality improvement: Iatrogenic venous air embolism in diagnostic imaging
Lockwood, P. and Breen, W. 2013. Patient safety and quality improvement: Iatrogenic venous air embolism in diagnostic imaging.
CT head reporting by radiographers: Findings of an accredited postgraduate programme
Lockwood, P. and Piper, K. 2013. CT head reporting by radiographers: Findings of an accredited postgraduate programme.
CT skull base & calvarium normal variant pitfalls
Lockwood, P. 2013. CT skull base & calvarium normal variant pitfalls.
CT head reporting by radiographers: results of an accredited postgraduate programme
Lockwood, P., Piper, K. and Pittock, L. 2015. CT head reporting by radiographers: results of an accredited postgraduate programme. Radiography. 21 (3), pp. e85-e89. https://doi.org/10.1016/j.radi.2014.12.001